McIlwain Tissue Chopper.

 
 
51350-mcilwain-l_web-705x538.jpg

Description.

This apparatus has been designed to prepare pieces of tissue for metabolic experiments and to be especially applicable to small and irregular specimens available at biopsy or from small organs. Its special features are to cause much less disturbance of cell structure than homogenisers or blenders, and to be applicable to fragments which would be difficult or impossible to cut by ordinary methods. It has been applied to liver, kidney and various parts of the central nervous system.

  • Minimises Cell damage

  • Tissue preparation for metabolic experiments

  • Ideal for cutting small irregular small organ or biopsy samples

Slices up to 1mm thick and cubes or prisms up to 1mm cross section can be prepared with the chopper in 30 seconds. Provision is made for stepless variation of slice thickness from zero to the maximum by means of an indexing clutch in conjunction with a micrometer head, the latter being calibrated in microns.
The action of the machine is as follows: a circular stainless steel table, on which the specimen is placed, is traversed automatically and at a speed which can be varied, from left to right of the machine. At the same time a chopping arm carrying a blade is raised and dropped by its own weight and the tension of a spring, at speeds varying from zero to over 200 strokes per minute. As the table carrying the tissue passes under the blade it is chopped into the required slices. If prisms are required the table is returned to the left hand side of the machine, turned through 45 degrees, and the process repeated. Cubes are formed by rotating the table 90 degrees. A safety limit switch is provided to prevent the machine overrunning and a quick return mechanism is fitted for the return of the table. Finish is in white powder coat.

Features

  • Stepless variation of slice thickness from zero to maximum.

  • The specimen table is traversed automatically from left to right at a variable speed.

  • Chopping Arm is raised and dropped at speeds of up to 200 strokes per minute.

  • Safety limit switch prevents overrunning and a quick return mechanism is fitted for the return of the table.

Specifications

  • Slice Thickness: maximum 1mm

  • Power Supply: 220-240 volts 50Hz or 115 volts 60HZ

  • Power Rating: 0.5 amp

  • Fuse: 3.15 amp anti surge (T) 5x20mm

  • Operating Conditions:

    • Indoor use

    • Altitude: < 2000 meters

    • Ambient Temperature: 5-40° C

    • Maximum RH: 80% at 31° C decreasing linearly to 50% RH at 40° C

    • Installation Category II

    • Pollution: degree 1 or 2 (IEC664)

  • Dimensions: 32 x 28 x 15 cms

  • Weight: 6.6kg (240 volt) 6.3 kg (115 volt)

McIlwain Tissue Chopper 220-240v 50Hz Product code: 21-805-02
McIlwain Tissue Chopper 110-115v 60Hz Product code: 21-800-03
Replacement plastic discs (pk/10) Product code: 21-810-10
Spare blades (pk/10) Product code: 21-815-10